Double-door high and low temperature and humidity test chamber
The double-door high and low temperature and humidity test chamber is used to test large components or complete machine samples for high and low temperature storage tests, high and low temperature cycle tests, temperature and humidity tests (alternating humidity and heat, constant humidity and heat), and high temperature and high humidity aging tests.
Category:
High and low temperature test chamber
Keyword:
Kexun
Mobile:18688631772 18925806499
Product Introduction
Product Specifications: Meets 80~1500L standard capacity and other non-standard customized models
Product Introduction: The double-door high and low temperature humidity test chamber is used to test large components or complete machine samples for high and low temperature storage tests, high and low temperature cycle tests, temperature and humidity tests (alternating humidity, constant humidity), and high temperature and high humidity aging tests.
Product Alias: Left and right opening high and low temperature humidity test chamber, double-door constant temperature and humidity chamber, large capacity double-door environmental test chamber, 1500L large temperature and humidity test chamber
The double-door high and low temperature humidity test chamber meets test standards such as MIL, STD, GB, GJB, JIS, JEDEC, and IEC. It can simulate the environment for high and low temperature, high and low temperature cycle, alternating humidity, constant temperature and humidity, high temperature and high humidity (double 85, double 95, double 98), and low temperature and low humidity reliability tests.
I. Application Fields: Semiconductor chips, scientific research institutions, quality inspection, new energy, optoelectronic communication, aerospace military industry, automotive industry, LCD display, medical and other technological industries
II. Test Standards Met:
1.1、GB/T 2423.1 Low temperature test method;
1.2、GB/T 2423.2 High temperature test method;
1.3、GB/T2423.34 Damp heat cycle test method;
1.4、IEC60068-2 Temperature and humidity test method;
1.5、GJB 150.3 High temperature test;
1.6、GJB 150.4 Low temperature test;
1.7、GJB 150.9 Damp heat test;
1.8、MIL-STD-202G-103B Humidity test;
1. Fluid simulation technology + innovative structural design, the equipment life is extended by more than 1/3 compared with traditional technology;
2. Scientific air duct design, temperature and humidity are stable and fast, and uniformity is better controlled;
3. New servo flow control technology, energy saving about 45% (long-term operation for more than 1 year without shutdown);
4. Extra-large test space, meeting the needs of small batch or large-scale machine testing, improving test efficiency;
5. Automatic load adjustment, stable performance;
6. Water saving (low water consumption for humidity, using full water discharge to ensure clean water and low failure rate);Technical Specifications
*Customizable non-standard models with any specifications*
|
Double-door high and low temperature (humidity) test chamber |
||
|
Model |
KS-HW-1500L A/B/D |
|
|
Working Volume (L) |
1500 L |
|
|
Inner Dimensions |
Width W * Depth D * Height H |
1500*1000*1000 |
|
External Dimensions |
Width W * Depth D * Height H |
1700*2000*2125 |
|
Single Window Size |
Width W * Height H |
290*420 |
|
Performance Indicators |
|
Temperature Performance Indicators |
|
Temperature Range |
A:-40~+150℃,B:-20~+150℃,D:-70~+150℃ |
|
|
Temperature Fluctuation |
≤±0.3℃ |
|
|
Temperature Uniformity |
≤1.5℃ |
|
|
Temperature Deviation |
≤±1.0℃ |
|
|
Heating Rate |
≥4℃/min average |
|
|
Cooling Rate |
≥1℃/min average |
|
|
|
Humidity Performance Indicators |
|
|
Humidity Range |
10%R.h~98%R.h |
|
|
Humidity Uniformity |
≤3.0%R.h |
|
|
Humidity Fluctuation |
≤±2.0%R.h |
|
|
Humidity Deviation |
±3.0%R.h |
|
|
Cabinet Material |
Inner Chamber |
SUS #304 Stainless Steel |
|
Outer Chamber |
Cold-rolled steel plate + double-sided electrostatic spraying process |
|
|
Insulation |
Ultra-fine glass wool |
|
|
Load |
According to customer samples |
|
|
Noise (dB) |
≤70 |
|
|
Wind Speed (m/s) |
≤1.7 |
|
|
Controller |
7-inch color touch screen controller (with pressure, temperature, and humidity sensors; with RS-485 or RS-232 communication; USB data baking functionality) |
|
|
Temperature and humidity time resolution |
Temperature: 0.01℃; Humidity: 0.1%; Time: 0.01min |
|
|
Cooling method |
Air cooling/Water cooling |
|
|
Refrigerant model |
Environmentally friendly refrigerants R404A, R23 |
|
|
Standard configuration |
Observation window (double-layer hollow tempered conductive glass) x1, test hole Φ100mm (located on the left) x1, sample rack 2 layers, in-chamber photographic lamp (LED light) x1, water supply tank x1, psychrometric band for wet bulb x1, small casters x4, power cord x1 |
|
|
Environmental requirements |
Air cooling environment temperature requirement +5~+30℃/Water cooling environment temperature requirement +5~+40℃ (water cooling requires a cooling tower or chiller) |
|
|
Power supply |
AC 380V 50/60Hz |
|
|
Weight (KG) |
850 |
|
|
Power (KW) |
9.5 |
|
|
Model description |
|
|