Rapid temperature change test chamber
The Rapid Temperature Change Test Chamber for Products (Environment Stress Screen Chamber) is suitable for environmental stress screening tests. By subjecting products to environmental stress screening, it accelerates the discovery of design flaws and improves product reliability.
Category:
Rapid temperature change test chamber
Keyword:
Kexun
Mobile:18688631772 18925806499
Product Introduction
Product Introduction: The product rapid temperature change test chamber (Environment Stress Screen Chamber) is suitable for environmental stress screening tests. By conducting environmental stress screening on products, it accelerates the discovery of design defects and improves product reliability.
Product Alias: Rapid temperature change test chamber, rapid temperature change and humidity test chamber, rapid temperature change test chamber, temperature change test chamber, temperature cycle test chamber, rapid heating test chamber, rapid temperature change chamber, temperature transient test chamber, high-speed temperature cycle test chamber
Rapid Temperature Change Test Chamber: Suitable for environmental stress screening (ESS) tests of electronic and electrical products, and temperature stress detection, temperature and humidity screening, reliability testing, performance testing, weathering testing, high and low temperature storage, etc., under conditions of rapid or gradual temperature changes. Commonly used temperature change rates are 5℃/min, 10℃/min, 15℃/min, 20℃/min, and 25℃/min.
1. The product meets the requirements of temperature setpoint change (linear) and average (non-linear) change.
2. Meets the temperature change rate requirement of 5~30℃/min.
3. Optional liquid nitrogen function, humidity function, anti-condensation function, etc. can be selected.
4. Using electronic expansion valve technology + innovative control system, the product is highly energy-efficient >45%.
I. Application Fields: Semiconductor chips, scientific research institutions, quality inspection, new energy, optoelectronic communication, aerospace military industry, automotive industry, LCD display, medical and other technology industries.
II. Operation Curve:
Features
Test Standards Met: GB/T 2423.1 Low temperature test method, GJB 150.3 High temperature test, GB/T 2423.2 High temperature test method, GJB 150.4 Low temperature test, GB/T2423.34 Humid heat cycle test method, GJB 150.9 Humid heat test, IEC60068-2 Temperature and humidity test method, MIL-STD-202G-103B Humidity test
*Customizable non-standard models with any specifications and parameters
Rapid Temperature Change (Humid Heat) Test Chamber |
||||||||||||||||||||||||||||||||||
|
|
KS-Without Humidity (Dry Heat) |
||||||||||||||||||||||||||||||||
Model Medal |
x: Linear temperature F: Non-Linear temperature |
KS-WB-150L-X/F |
KS-WB-225L-X/F | KS--WB-512L-X/F | KS-WB-1000L-X/F | KS-WB-2400L-X/F | KS-WB-3400L-X/F | |||||||||||||||||||||||||||
KSH-With Temperature Bottom (Humid Heat) |
||||||||||||||||||||||||||||||||||
KSH-WB-150L-X/F |
KSH-WB-225L-X/F |
KSH-WB-512L-X/F |
KSH-WB-1000L-X/F |
KSH-WB-2400L-X/F |
KSH-WB-3400L-X/F |
|||||||||||||||||||||||||||||
Internal Volume (L) |
150L |
225L |
512L |
1000L |
2400L |
3400L |
||||||||||||||||||||||||||||
Internal Dimensions (mm) |
Width |
500 |
600 |
800 |
1000 |
1300 |
1500 |
|||||||||||||||||||||||||||
Depth |
500 |
500 |
800 |
1000 |
1400 |
1500 |
||||||||||||||||||||||||||||
Height |
600 |
750 |
800 |
1000 |
1300 |
1500 |
||||||||||||||||||||||||||||
Heating and Cooling Rate (℃/min) |
5 |
10 |
15 |
20 |
25 |
5 |
10 |
15 |
20 |
25 |
5 |
10 |
15 |
20 |
25 |
5 |
10 |
15 |
20 |
25 |
5 |
10 |
15 |
20 |
25 |
5 |
10 |
15 |
20 |
25 |
||||
Performance Indicators |
|
Humidity Performance Indicators |
||||||||||||||||||||||||||||||||
Temperature Range |
-70℃-+180℃ |
|||||||||||||||||||||||||||||||||
Rapid Temperature Change Range |
A: 40~+125℃, C: -55-+125℃ |
|||||||||||||||||||||||||||||||||
Temperature Uniformity |
≤2.0℃ |
|||||||||||||||||||||||||||||||||
Temperature Fluctuation |
≤±0.5℃ |
|||||||||||||||||||||||||||||||||
Temperature Deviation |
≤±2.0℃ |
|||||||||||||||||||||||||||||||||
|
Humidity Performance Indicators |
|||||||||||||||||||||||||||||||||
Humidity Range |
10%RH~98%RH, (5%~98%RH; special conditions are customized) |
|||||||||||||||||||||||||||||||||
Humidity Uniformity |
≤3.0%RH |
|||||||||||||||||||||||||||||||||
Humidity Fluctuation |
≤±2.0%RH |
|||||||||||||||||||||||||||||||||
Humidity Deviation |
±3.0%RH(>75%RH); ±5.0%RH(≤75%RH) |
|||||||||||||||||||||||||||||||||
Cabinet Material |
Inner Chamber |
SUS#304 Stainless Steel |
||||||||||||||||||||||||||||||||
Outer Chamber |
Cold-rolled sheet + double-sided electrostatic spraying process, Kexun standard blue-gray |
|||||||||||||||||||||||||||||||||
Insulation Layer |
Ultra-fine glass wool |
|||||||||||||||||||||||||||||||||
Viewing Window Size (mm) |
Width W x Height H |
210*300 |
280*350 |
360*510 |
360*510 |
360*510 |
360*510 |
|||||||||||||||||||||||||||
Noise |
≤75(dB) |
|||||||||||||||||||||||||||||||||
Cooling Method |
Water cooling (water temperature 7~28℃, water pressure 0.1-0.4Mpa) |
|||||||||||||||||||||||||||||||||
Environmental Requirements |
Ambient temperature: 5~35℃, relative humidity: ≤85%RH, no strong vibration, electromagnetic radiation shielding, no dust or corrosive substances around |
|||||||||||||||||||||||||||||||||
Standard Configuration |
Observation window (double-layer hollow conductive glass) 1, test hole φ50/φ100mm (located on the left) 1, sample rack 2 layers. In-chamber photographic lamp (LED lamp) 1. Water tank 1, psychrometer with meteorological tape 1, small casters 4, power cord 1, instruction manual 1 |
|||||||||||||||||||||||||||||||||
Power Supply |
Three-phase four-wire + ground wire (3/N/PE), AC380V±10%, 50Hz±1, ground resistance less than 40 |
|||||||||||||||||||||||||||||||||
Model Description |
|