DQY High-Altitude Low-Pressure Test Chamber
The high-altitude, low-pressure test chamber is used to determine the adaptability and reliability of components, materials, instruments, equipment, or other products in (high-altitude, low-oxygen, high-cold, plateau, etc.) environments. It can also meet the requirements of high and low temperature/constant temperature and humidity tests, and detect the parameter performance of various materials after changes in extreme environments such as heat resistance, cold resistance, dryness resistance, and humidity resistance.
Category:
High and low temperature test chamber
Keyword:
Kexun
Mobile:18688631772 18925806499
Product Introduction
Product Specifications: Inner box meets 150~3300L standard volume and other non-standard customized models
Product Use: The product high-altitude low-pressure test chamber is used to determine the adaptability and reliability of components, materials, instruments, equipment, or other products in (high altitude, low oxygen, high cold, plateau, etc.) environments. It can also meet the tests of high and low temperature/constant temperature and humidity changes, and detect the parameter performance of various materials after changes in heat resistance, cold resistance, dryness resistance, and humidity resistance in extreme environments.
Product Alias: High and low temperature low pressure test chamber, high and low temperature humidity low pressure test chamber, high altitude environmental test chamber, plateau low pressure test machine.
The KSKexun high-altitude low-pressure test chamber simulates the changes in temperature and altitude to evaluate the use and storage of electronic and electrical equipment in (high altitude, low oxygen, high cold, plateau, etc.) environments. It is widely used in the reliability testing of aerospace products, information communication, instrumentation, electrical engineering, and mechanical products.
I. Test Standards Met:
GB/T 2423.1-2008 Test A: Low temperature test method;
GJB150.1A-2A-3A-4A-9A-24A-2009 Military equipment laboratory environmental test methods.
GB/T 2423.21-2008 Test M: Low pressure test method;
GB/T 2423.25-2008 Test Z/AM: Low temperature/low pressure combined test.
GB/T 2423.26-2008 Test Z/BM: High temperature/low pressure combined test;
GB/T 2423.2-2008 Test B: High temperature test method.
GJB360A-1996 Method 105: Low pressure test;
GJB 367A-2001 "General Specifications for Military Communication Equipment" Appendix A General Environmental Test Methods.
II. Equivalent Altitude as shown in the table below ((from GB/T 2423.21-2008))
|
Set Pressure (KPa) |
1 |
2 |
4 |
8 |
15 |
25 |
40 |
55 |
70 |
84 |
|
Corresponding Altitude (m) |
31200 |
26600 |
22100 |
17600 |
13600 |
10400 |
7200 |
4850 |
3000 |
1000 |
1. Advanced pressure control system, small fluctuation, high stability;
2. Single compressor refrigeration system, high efficiency, low energy consumption, low operating cost;
3. Compact structure design, light weight, small footprint, easy transportation and installation;
4. Performance indicators exceed national (military) standard quality requirements;
5. Using electronic expansion valve technology and innovative control system, product energy efficiency >30%; Specifications
*Customizable non-standard models with any specification parameters
|
High Altitude (Altitude) Low Pressure Test Chamber |
||||||
|
Model |
KS-DQY-225L |
KS-DQY-408L |
KS-DQY-1000L |
KS-DQY-1700L |
KS-DQY-3300L |
|
|
Internal Volume (L) |
225 L |
408 L |
1000 L |
1700 L |
3300 L |
|
|
Inner Chamber Dimensions |
Width W |
600 |
800 |
1000 |
1200 |
1500 |
|
Depth D |
500 |
650 |
1000 |
1200 |
1500 |
|
|
Height H |
750 |
850 |
1000 |
1200 |
1500 |
|
|
Performance Indicators |
|
Temperature Performance Indicators |
||||
|
Temperature Range |
A: -40~+150℃, B: -20~+150℃, D: -70~+150℃ |
|||||
|
Temperature Fluctuation |
≤±0.3℃ (≤±0.5℃, according to GB/T5170-1996) |
|||||
|
Temperature Uniformity |
≤2.0℃ (at normal pressure, no load) |
|||||
|
Temperature Deviation |
≤±2℃ (at normal pressure, no load) |
|||||
|
Heating Rate |
Average 3℃/min (at normal pressure, no load) |
|||||
|
Cooling Rate |
Average 2℃/min (at normal pressure, no load); |
|||||
|
|
Humidity Performance Indicators (Optional) |
|||||
|
Humidity Range |
20%R.h~98%R.h (other special conditions require customization) |
|||||
|
Humidity Uniformity |
≤3.0%RH |
|||||
|
Humidity Fluctuation |
≤±2.0%RH |
|||||
|
Humidity Deviation |
±3.0%RH (>75%RH), ±5.0%RH (≤75%RH) |
|||||
|
|
Pressure Performance Indicators |
|||||
|
Pressure Range |
Normal pressure~0.5KPa |
|||||
|
Pressure Deviation |
Normal pressure~40kPa: ≤±1.8kPa; 40kPa~4kPa: ≤±4.5% kPa; 4kPa~1kPa: ≤±0.1kPa |
|||||
|
Pressure Drop Time |
Normal pressure→0.5kPa ≤ 45min (no load) |
|||||
|
Pressure recovery rate |
≤10kPa/min |
|||||
|
Cabinet material |
Inner box |
SUS #304 stainless steel |
||||
|
Outer box |
Cold-rolled steel plate + double-sided electrostatic spraying process |
|||||
|
Insulation |
Ultra-fine glass wool |
|||||
|
Pressurization method |
Internal pressurization |
|||||
|
Noise |
≤70 (dB) |
|||||
|
Controller |
7-inch color touch screen controller (with pressure, temperature and humidity, with RS-485 or RS-232 communication, USB data baking) |
|||||
|
Temperature, humidity and time resolution |
Temperature: 0.01℃; Humidity: 0.1%; Time: 0.01min |
|||||
|
Cooling method |
Water cooling, air cooling |
|||||
|
Refrigerant model |
Environmentally friendly refrigerant R404A, R23 |
|||||
|
Environmental requirements |
Air-cooled ambient temperature requirement +5~+30℃/Water-cooled ambient temperature requirement +5~+40℃ (water cooling requires a water tower or chiller) |
|||||
|
Power supply |
AC380V±10%, 50Hz±1 three-phase four-wire + ground wire (3/N/PE), ground resistance less than 4Ω |
|||||
|
Model description |
|
|||||