Product Center

Contact Us

Contact person: Mr. Wang




No. 1, Building 1, Zhenrong Road, Wusha, Chang'an Town, Dongguan City, Guangdong Province


+
  • 图片 70.jpg
  • 图片 71.jpg
  • 图片 72.jpg

DQY High and Low Temperature (Low Pressure) Test Chamber

Evaluate the use and storage of electronic and electrical equipment under single or combined conditions simulating temperature and altitude changes (high altitude, low oxygen, high cold, plateau, etc.). Widely used in reliability testing of aerospace products, information communication, instrumentation, electrical, and mechanical products.

Category:

High and low temperature test chamber

Keyword:

Kexun



Product Introduction

Product Specifications: Inner box meets 225~3300L standard volume and other non-standard customized models

Product Introduction: Evaluates the use and storage of products, components, and materials in electronic and electrical equipment under single or combined conditions simulating temperature and altitude changes (high altitude, low oxygen, high cold, plateau, etc.). Widely used in the reliability testing of aerospace products, information communication, instrumentation, electrical engineering, and mechanical products.

Product Alias: Altitude Temperature Environmental Test Chamber, High Altitude Low Pressure Test Chamber, Low Pressure Chamber, Plateau Environmental Test Chamber, High Temperature High Cold Altitude Test Chamber, Low Temperature Low Pressure Environmental Chamber, Temperature Humidity High Comprehensive Test Chamber

I. Test Standards Met:

GB/T 2423.1-2008 Test A: Low Temperature Test Method;

GJB150.1A-2A-3A-4A-9A-24A-2009 Military Equipment Laboratory Environmental Test Methods.

GB/T 2423.21-2008 Test M: Low Pressure Test Method;

GB/T 2423.25-2008 Test Z/AM: Low Temperature/Low Pressure Combined Test.

GB/T 2423.26-2008 Test Z/BM: High Temperature/Low Pressure Combined Test;

GB/T 2423.2-2008 Test B: High Temperature Test Method.

GJB360A-1996 Method 105: Low Pressure Test;

GJB 367A-2001 "General Specifications for Military Communication Equipment" Appendix A General Environmental Test Methods.

II. Equivalent Altitude as shown in the table below ((extracted from GB/T 2423.21-2008))

Set Pressure (KPa)

1

2

4

8

15

25

40

55

70

84

Corresponding Altitude (m)

31200

26600

22100

17600

13600

10400

7200

4850

3000

1000

1. High-efficiency bipolar vacuum pump pumping unit + DPT electronic pressure regulating valve, balance compensation type pressure control;

2. Using centrifugal fan, magnetic fluid drive method internal circulation;

3. International brand (German Leybold) bipolar rotary vane vacuum pump;

4. Special design explosion-proof armored electric heating tube;

5. Special design mesh distributed evaporator;

6. High thermal radiation characteristic mesh inner box wall plate;

Technical Specifications

*Customizable arbitrary specification parameters non-standard models

High and Low Temperature (Low Pressure) Test Chamber

Model

KS-DQY-225L-A/B/D

KS-DQY-408L-A/B/D

KS-DQY-1000L-A/B/D

KS-DQY-1700L-A/B/D

KS-DQY-3300L-A/B/D

Internal Volume (L)

225 L

408 L

1000 L

1700 L

3300 L

Inner Chamber Dimensions
(mm)

Width W

600

800

1000

1200

1500

Depth D

500

650

1000

1200

1500

Height H

750

850

1000

1200

1500

Performance Indicators

 

Temperature Performance Indicators

Temperature Range

A: -40~+150℃, B: -20~+150℃, D: -70~+150℃

Temperature Fluctuation

≤±0.3℃ (≤±0.5℃, according to GB/T5170-1996)

Temperature Uniformity

≤2.0℃ (normal pressure, no-load)

Temperature Deviation

≤±2℃ (normal pressure, no-load)

Heating Rate

Average 3℃/min (normal pressure, no-load)
(Temperature rate 4~25℃/minute linear/non-linear is a special condition that needs to be customized)

Cooling Rate

Average 2℃/min (normal pressure, no-load);
(Temperature rate 2~25℃/minute linear/non-linear is a special condition that needs to be customized)

 

Humidity Performance Indicators (Optional)

Humidity Range

20%R.h~98%R.h (other special conditions need to be customized)

Humidity Uniformity

≤3.0%RH

Humidity Fluctuation

≤±2.0%RH

Humidity Deviation

±3.0%RH (>75%RH), ±5.0%RH (≤75%RH)

 

Pressure Performance Indicators

Pressure Range

Normal pressure~0.5KPa

Pressure Deviation

Normal pressure~40kPa: ≤±1.8kPa; 40kPa~4kPa: ≤±4.5% kPa; 4kPa~1kPa: ≤±0.1kPa

Pressure Drop Time

Normal pressure→0.5kPa ≤ 45min (no-load)

Pressure Recovery Rate

≤10kPa/min

Cabinet Material

Inner box

SUS #304 stainless steel

Outer box

Cold-rolled steel plate + double-sided electrostatic spraying process

Insulation

Ultrafine glass wool

Pressure bearing method

Internal pressure bearing

Noise

≤70 (dB)

Controller

7-inch color touch screen controller (with pressure, temperature and humidity, with RS-485 or RS-232 communication, USB data baking)

Temperature, humidity and time resolution

Temperature: 0.01℃; Humidity: 0.1%; Time: 0.01min

Cooling method

Water cooling, air cooling

Refrigerant model

Environmentally friendly refrigerant R404A, R23

Environmental requirements

Air cooling environment temperature requirement +5~+30℃/Water cooling environment temperature requirement +5~+40℃ (Water cooling requires a water tower or chiller)

Power supply

AC380V±10%, 50Hz±1 three-phase four-wire + ground wire (3/N/PE), ground resistance less than 4Ω

Model description